EDL’s Surface Transfer Standard calibrators are perfect for accurate surface calibrations of surface sensors. The STS-SC1 & STS-SC2 were designed for ease of use, portability, and rugged durability. Both devices use a Eurotherm Controller with ITools for easy automation of your temperature measurement process.
Calibrations of surface sensors are easily and cost effectively performed in-house and laboratory with the STS line. By eliminating the need to send sensors out to costly calibration labs, the STS will save you money by extending the lifespan of sensors and increasing the total accuracy of your measurements.
A novel method was used to link the STS to a traceable standard. A single sensor during calibration provides no measure of the heat flux within the block. Thermal interfaces and differences in thermal conductivity between the block and sensor also prevent traceable calibration. In the STS, a thermocouple is embedded in a precisely machined bore to establish a consistent and precise fit between the thermocouple and block. This allows an external pyrometer, or temperature measuring instrument, to be used during the calibration to verify the true temperature. Calibrating surface sensors in-house is simplified with this product line, requiring only that measurements are made of the reference thermocouple together with the sensor being tested. made of the reference thermocouple together with the sensor being tested.
Standard Features
• Portable & Rugged Industrial Design
• Surface Thermocouple Connection Outputs For Added Accuracy & Reliability
• Expanded Uncertainties Of ± 1.0 °C Achieved For Entire Operating Temperature Range
• Eurotherm Controller With ITools For Easy Automation
• Initial & Annual Re-Calibration Performed In Our Calibration Laboratory
• 12-Month Warranty
STS-SC1 & STS-SC2 > Surface Temperature Calibration
Note: Typical lead time ARO on Metrology Calibration products is 6 to 10 weeks. Expediting is avaiable apon request, at an additional cost, please call sales at 084.8.222.16361 for more information.
Specifications | ||
Part Number | STS-SC1 | STS-SC2 |
Range | Ambient to 650° C | Ambient to 315° C |
Size | 30.4cm x 25.4cm x 34.6cm (12" x 10" x 14") |
|
Weight | 11 kg (25 lbs) | 9 kg (22 lbs) |
Resolution |
0.1° C | |
Accuracy | ±2.0° C | |
Stability | ±0.2° C (Over 30 min. period) |
±0.1° C (Over 30 min. period) |
Uniformity | ±0.7° C | ±0.5° C |
Fast Cool | 200° C in ~30 min. | 200° C in ~40 min. |
Warm-up Time | 100° C in ~5 min. 300° C in ~15 min. |
100° C in ~10 min. 300° C in ~20 min. |
Ordering Information
In order to determine your part number refer to the chart below. The letters and numbers highlighted in red create the part number. Read from left to right in the column rows to create the part number (ex. STS-SC1). Click here for additional ordering information.
Model | Range |
STS-SC1 Surface Transfer Standard
|
Ambient to 650° C (High Temperature) |
STS-SC2 Surface Transfer Standard
|
Ambient to 315° C |
… [Trackback]
[…] Read More Info here to that Topic: ino.com.vn/khac-do-luong-kiem-tra/edl-surface-transfer-standard-modelsts-sc1-sts-sc2/ […]
… [Trackback]
[…] Info on that Topic: ino.com.vn/khac-do-luong-kiem-tra/edl-surface-transfer-standard-modelsts-sc1-sts-sc2/ […]
… [Trackback]
[…] Read More Information here to that Topic: ino.com.vn/khac-do-luong-kiem-tra/edl-surface-transfer-standard-modelsts-sc1-sts-sc2/ […]
… [Trackback]
[…] Info on that Topic: ino.com.vn/khac-do-luong-kiem-tra/edl-surface-transfer-standard-modelsts-sc1-sts-sc2/ […]
… [Trackback]
[…] Read More to that Topic: ino.com.vn/khac-do-luong-kiem-tra/edl-surface-transfer-standard-modelsts-sc1-sts-sc2/ […]