CALYS 80P is a field pressure calibrator within CALYS range with high accuracy of 0.01%. CALYS are the perfect tools for advanced process maintenance and use on test bench in non-hazardous areas. Suitable for all field and lab measurements, CALYS 80P is dedicated to simultaneous measurement and generation over two isolated channels of pressure and process signals. It can also measure resistance signals and resistive probes.
Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…) and audit trails, CALYS series makes advanced data exploitation and full data traceability easier. CALYS can store and recall up to 10 complete instrument configurations with manual or automatic recall for easy and quick work in the field and more than 1,500 measurements with date and time.
The dual channel display allows a simultaneous indication of the measured and the simulated values. The graphic mode allows the trend to be displayed.
Specifications and performances in temperature @23°C ±5°C
Uncertainty is given in % of reading (CALYS 80P display) + fixed value.
Resistive probes: Measurement and simulation
Sensor | Range (Input and Output) | Resolution | Accuracy / 1 year |
Pt100 (= IEC) | -200°C to +850°C | 0.01°C | 0.01% RDG + 0.05°C |
Pt100 (= OIML) | -200°C to +850°C | 0.01°C | 0.01% RDG + 0.05°C |
Pt100 (= 3926) | -200°C to +850°C | 0.01°C | 0.01% RDG + 0.05°C |
Pt100 (= 3902) | -200°C to +650°C | 0.01°C | 0.01% RDG + 0.05°C |
Pt100 (= JIS SAMA) | -200°C to +600°C | 0.01°C | 0.01% RDG + 0.05°C |
Pt200 (= 3851) | -200°C to +850°C | 0.1°C | 0.01% RDG + 0.15°C |
Pt500 (= 3851) | -200°C to +850°C | 0.1°C | 0.01% RDG + 0.1°C |
Pt1000 (= IEC) | -200°C to +850°C | 0.1°C | 0.01% RDG + 0.1°C |
Pt1000 (= OIML) | -200°C to +850°C | 0.1°C | 0.01% RDG + 0.1°C |
Cu10 (= 427) | -70°C to +150°C | 0.1°C | 0.01% RDG + 0.4°C |
Cu100 (= 428) | -180°C to +150°C | 0.1°C | 0.01% RDG + 0.05°C |
Ni100 (= 618) | -60°C to +180°C | 0.1°C | 0.01% RDG + 0.05°C |
Ni120 (= 672) | 0°C to +150°C | 0.1°C | 0.01% RDG + 0.05°C |
Connections: 2, 3 and 4 wires
Rtd measurement excitation current: 0.2 mA
Rtd cable compensation: up to 100 mΩ (for each wire)
Rtd cable compensation error (Pt100): ±0.005°C/ of total wire
Maximum load resistance: 600 Ω @ 20 mA
Specifications and performances in pressure @23°C ±5°C
Pressure
Pressure media: AISI 316 SS compatible fluids (water, gas, and oil)
Temperature compensation: Automatic with built-in calibration matrix
Engineering units: mbar, bar, hPa, kPa, Mpa, kg/cm2, kg/m2, psi, mmH2O, cmH2O, mH2O, Torr, atm, lb/ft2, inH2O, FTH2O, mmHg, cmHg, mHg, inHg
Accuracy: accuracies given for 1 year, include non linearity, hysteresis, and repeatability.
Temperature coefficient inside the temperature compensated range: ±0.002% RDG /°C
Compensation range: 0°C to 45°C
Measurement by internal pressure module
Reference | Range | Resolution |
IPM000100G | -100 / 100 mbar Gauge | 0.001 mbar |
IPM000500G | -500 / 500 mbar Gauge | 0.01 mbar |
IPM002000G | -0.95 / 2 bar Gauge | 0.01 mbar |
IPM007000G | -0.95 / 7 bar Gauge | 0.1 mbar |
IPM020000G | -0.95 / 20 bar Gauge | 0.1 mbar |
IPM002000A | 2 bar Absolute | 0.01 mbar |
IPM020000A | 20 bar Absolute | 0.1 mbar |
Accuracy: ±0.025% F.S.
Overpressure: 125% F.S.
Port: female, 1/8" BSP
… [Trackback]
[…] Find More on that Topic: ino.com.vn/aoip/aoip-field-pressure-calibrator-with-high-accuracy-0-01-modelcalys-80p/ […]
… [Trackback]
[…] Read More here on that Topic: ino.com.vn/aoip/aoip-field-pressure-calibrator-with-high-accuracy-0-01-modelcalys-80p/ […]