- MX100 Style S3 New!
- 30-ch input module
- Windows Vista compatible
- Alarm levels increased from 2 to 4
- Multi-Channel Capability, High Isolation Voltage
- Multi-Interval Measurement
- Flexible System Configuration
- Versatile PC-Based Software Options
- Easy Software Setup
- No Re-Wiring between Measurements
- All-in-one "Quick Start Packages" with 10, 20, or 30 Channels
Features
High-Speed Multi-Channel Capability
The table below provides reference information on the maximum number of measurable channels in relevant measurement intervals (when using Yokogawa's proprietary software):
The relationship between measurement intervals and the number of channels largely depends on the performance of the PC. The actual performance may differ from that shown in the table.
<PC System Requirements>
CPU: Pentium 4 3.2 GHz; Memory: 1 GB; OS: Windows 2000; Hard disk: 160 GB
Communication interface: Ethernet 100Base-TX
Multi-Interval Measurement
Three types of measurement intervals can be used in a single system in various combinations. Measurement intervals are specified for each module. It is also possible to set different types of input modules to the same measurement interval or to set the same type of input modules to different measurement intervals. Twelve measurement intervals are available as userselectable options: 10 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s and 60 s. Please note, however, that measurement intervals of 10/50 ms cannot be set to the Medium-speed Universal Input Module (MX110-UNV-M10).
Examples of Setting Measurement Intervals
Multi-interval systems can be flexibly built for each module as shown in the figure below. Three colors are used to indicate the three types of measurement intervals. |
Monitor Window Using Yokogawa's Proprietary PC Software
The window shows measurement values by measurement intervals. Measurement can be performed while confirming rapidly-changing signals and slowly-changing signals simultaneously. |
A: Measurement of sine waves at 0.5 Hz with a measurement interval of 100 ms
B: Measurement of sine waves at 8 Hz with a measurement interval of 10 ms
C: Measurement of sine waves at 1 Hz with a measurement interval of 50 ms
High Withstand Voltage (Reinforced Insulation)
Reinforced (double) insulation has been provided between the input terminals and the case to achieve a high withstand voltage of 600 Vrms/VDC (continuous). As shown below, each battery voltage can be measured even if the batteries are stacked to increase common-mode voltages.
Noise Rejection (MX110-UNV-H04)
MX input modules employ integrating A/D converters with superior noise rejection performance. In particular, the MX110-UNV-H04 Four-Channel High-Speed Universal Input Module incorporates an integrating A/D converter for each input channel, thereby exhibiting strong noise rejection performance in high-speed measurement. Moreover, the digital filtering function is effective on noises that cannot be fully rejected even with integrating A/D converters. Time constants can be selected from 5 to 100 measurement interval times. Therefore, filtering can be applied according to the noise to be rejected.
Example of Noise Rejection
When the noise source in the left figure was applied to the MX, the result in the right figure was obtained. High-speed measurement can be stably performed even in noisy environments.
Noise source applied to the MX A noise signal was applied between H and L (in normal mode). |
Output window of MX100 for noise inputs Module used: MX110-UNV-H04 Four-Channel High-Speed Universal Input Module Measurement with the range of 2 V (-2 V to +2 V), measurement interval of 50 ms, integral time “Auto” and filter “OFF.” Example of noise rejection |
System Configuration
Connection between a single MX unit and a single PC (measurement of 24 channels/10 ms or 60 channels/100 ms)
One MX unit enables data acquisition on up to 24 channels at a measurement interval of 10 ms, or up to 60 channels at a measurement interval of 100 ms (six modules installed in both cases).
Setting up the System The MX and PC are connected via a hub and a straight Ethernet cable. Alternatively, they can be connected using a cross cable alone, without using the hub |
|
Input signal: DC voltageThermocuple, RTD,Digital input (Mixed input possible) |
PC Software In a one-to-one connection, the MX100 standard software included with the main module facilitates easy building of data acquisition environments. One MX unit consists of a combination of a main module, input/output modules, and a base plate. This configuration centers on the main module. |
Output signal: Digital output |
Connection between multiple MX units and a single PC
For measurement of up to 1,200 channels and the use of MXLOGGER
Example of a system with multiple MX units:
Setting up the System The MX units and the PC are connected by means of a hub and straight Ethernet cables. PC Software Please note that the shortest measurement interval of MXLOGGER depends on the system's total number of channels and the PC's performance. For guidelines on measurement intervals and the number of channels, please visit…. If users wish to create their own data acquisition software, there is no limit to the number of MX units that can be integrated with a single PC software element. Please note, however, that if conditions with heavy loads (such as short measurement intervals with many units) are set to the PC and its software, data will not be acquired in time and some data may be lost. |
Hardware Configuration
ModuleThe MX is designed to enable desired measurement environments by combining three elements: the main module, input/output modules, and a base plate. The assembled unit can be utilized on the desktop as-is (the modules have legs). Please note that DIN rails are used for rack-mounting. |
Main Module (MX100)
The main module is the engine that controls data acquisition. It is equipped with a power supply, an Ethernet port, and a CompactFlash card slot. One main module can accommodate up to six input/output modules. The user can choose up to six moduels of any type.
Ethernet port: For communication with the PC. Automatic recognition of 100Base-TX or 10Base-T CF (CompactFlash) Card Slot: |
Data Backup Time by CF Card Size (Note: Minutes/hours/days are approximate.)
|
Input Modules
Name | Model | Number of channels | Shortest measurement interval | Description |
30 ch general purpose input module with 500 ms scan speed | MX110-VTDL30 | 30 | 500 ms | Clamp terminal, DCV/TC/DI |
MX110-VTDL30/H3 | 30 | 500 ms | M3 screw terminal, DCV/TC/DI |
Name | Model | Number of channels | Shortest measurement interval | Description |
Universal Input Modules | MX110-UNV-H04 | 4 | 10 ms | DC voltage, thermocouple, 3-wire RTD, DI (non-voltage contact, Level (5 V logic)). Mixed input allowed. |
MX110-UNV-M10 | 10 | 100 ms | DC voltage, thermocouple, 3-wire RTD, DI (non-voltage contact, Level (5 V logic)). Mixed input allowed. | |
4-Wire RTD and Resistance Input Module | MX110-V4R-M06 | 6 | 100 ms | DC voltage, 4-wire RTD, 4-wire resistance, DI (non-voltage contact, Level (5 V logic)). Mixed input allowed. |
Strain Input Modules | MX112-B12-M04 | 4 | 100 ms | Built-in bridge resistance of 120 ohm |
MX112-B35-M04 | Built-in bridge resistance of 350 ohm | |||
MX112-NDI-M04 | For connection with an external bridge head and strain gauge type sensor (NDIS connector)) | |||
5 V Digital Input Module | MX115-D05-H10 | 10 | 10 ms | Non-voltage contact, open collector, and Level (5 V logic). Mixed input allowed. |
24 V Digital Input Module | MX115-D24-H10 | 10 | 10 ms | Level (24 V logic), Vth = 12 V |
Output Modules
Name | Model | Number of channels | Ouput update interval | Description |
Analog Output Module | MX120-VAO-M08 | 8 | 100 ms | Allows mixed voltage (±10 V) and current (4-20 mA) output |
PWM Output Module | MX120-PWM-M08 | 8 | 100 ms | Pulse width modulation output module |
Digital Output Module | MX125-MKC-M10 | 10 | 100 ms | “A” contact (SPST) |
Base Plate
Base plates available for all configurations, from 1 to 6 input/output modules.
When used for the MW100/MX100, you must replace the attachment with the one that comes standard with the MW100/MX100.
Accessories
Connector Covers Connector covers for open slots AC Adapter (772075) |
Accessories (Removable Terminal)
All input/output terminals are removable except for those of the MX112-NDI-M04. If multiple terminals are prepared ahead of time, no re-wiring is needed between measurements. |
Model | Description |
772061 | Screw (M4) terminal block (RJC (reference junction compensation) included). For use in combination with 772062. Compatible with MX110-UNV-M10, MX115-D05-H10, and MX115-D24-H10. |
772062 | Connection cable between input modules and the screw terminal block. Compatible with MX110-UNV-M10, MX115-D05-H10, and MX115-D24-H10. |
772063 | Plate with clamp terminals (RJC included). Compatible with MX110-UNV-M10, MX115-D05-H10, and MX115-D24-H10. |
772064 | Clamp terminals. Compatible with MX110-UNV-H04. |
772065 | Clamp terminals. Compatible with MX120-VAO-M08, MX120-PWM-M08, and MX125-MKC-M10. |
772067 | Plate with clamp terminals. Compatible with MX110-V4R-M06. |
772068 | Plate with clamp terminals (Built-in bridge resistance of 120 ohm). Compatible with MX112-Bxx-M04. |
772069 | Plate with clamp terminals (Built-in bridge resistance of 350 ohm). Compatible with MX112-Bxx-M04. |
772080 | Plate with screw (M3) terminal (RJC included). Compatible with MX110-UNV-M10, MX115-D05-H10, and MX115-D24-H10. |
772081 | Plate with clamp terminal for current with 10 ohm built in bridge resistance, applies to MX110-UNV-M10 |